Rectilinear Ion Trap:  Concepts, Calculations, and Analytical Performance of a New Mass Analyzer

Zheng Ouyang, Guangxiang Wu, Yishu Song, Hongyan Li, Wolfgang R. Plass, and R. Graham Cooks*
Department of Chemistry, Purdue University, West Lafayette, Indiana 47907-2084, and II. Physikalisches Institut, Justus-Liebig-Universität Giessen, Heinrich-Buff-Ring 16, 35392 Giessen, Germany
Anal. Chem., 2004, 76 (16), pp 4595–4605
DOI: 10.1021/ac049420n
Publication Date (Web): July 16, 2004
Copyright © 2004 American Chemical Society

Abstract

A mass analyzer based on a rectilinear geometry ion trap (RIT) has been built, and its performance has been characterized. Design concepts for this type of ion trap are delineated with emphasis on the effects of electrode geometry on the calculated electric field. The Mathieu stability region was mapped experimentally. The instrument can be operated using mass-selective instability scans in both the boundary and resonance ejection versions. Comparisons of performance between different versions of the device having different dimensions allowed selection of an optimized geometry with an appropriate distribution of higher-order electric fields. Comparisons made under the same conditions between the performance of a conventional cylindrical ion trap and a RIT of 4 times greater volume show an improvement of 40 times in the signal-to-noise ratio resulting from the higher ion trapping capacity of the RIT. The demonstrated capabilities of the RIT include tandem mass spectrometry, a mass resolution in excess of 1000, and a mass/charge range of 650 Th, all in a simple structure that is only 3.5 cm3 in internal volume.

Citing Articles

View all 69 citing articles

Citation data is made available by participants in CrossRef's Cited-by Linking service. For a more comprehensive list of citations to this article, users are encouraged to perform a search in SciFinder.

This article has been cited by 19 ACS Journal articles (5 most recent appear below).

  • Cover Image

    Coaxial Ion Trap Mass Spectrometer: Concentric Toroidal and Quadrupolar Trapping Regions

    Ying Peng , Brett Hansen , Hannah Quist , Zhiping Zhang , Miao Wang , A. R. Hawkins , and Daniel E Austin
    Analytical Chemistry 0 (ja),
    • Coaxial Ion Trap Mass Spectrometer: Concentric Toroidal and Quadrupolar Trapping Regions

      Ying Peng , Brett Hansen , Hannah Quist , Zhiping Zhang , Miao Wang , A. R. Hawkins , and Daniel E Austin
      Analytical Chemistry 0 (ja),

      We present the design and results for a new radio-frequency ion trap mass analyzer, the coaxial ion trap, in which both toroidal and quadrupolar trapping regions are created simultaneously. The device is composed of two parallel ceramic plates, the ...

  • Cover Image

    Sampling Wand for an Ion Trap Mass Spectrometer

    Keyong Hou, Wei Xu, Jian Xu, R. Graham Cooks, and Zheng Ouyang
    Analytical Chemistry2011 83 (5), 1857-1861
    • Sampling Wand for an Ion Trap Mass Spectrometer

      Keyong Hou, Wei Xu, Jian Xu, R. Graham Cooks, and Zheng Ouyang
      Analytical Chemistry2011 83 (5), 1857-1861

      A new sampling wand concept for ion trap mass spectrometers equipped with discontinuous atmospheric pressure interfaces (DAPI) has been implemented. The ion trap/DAPI combination facilitates the operation of miniature mass spectrometers equipped with ...

  • Cover Image

    Miniature Mass Spectrometer Systems Based on a Microengineered Quadrupole Filter

    Andrew Malcolm, Steven Wright, Richard R. A. Syms, Neil Dash, Marc-André Schwab and Alan Finlay
    Analytical Chemistry2010 82 (5), 1751-1758
    • Miniature Mass Spectrometer Systems Based on a Microengineered Quadrupole Filter

      Andrew Malcolm, Steven Wright, Richard R. A. Syms, Neil Dash, Marc-André Schwab and Alan Finlay
      Analytical Chemistry2010 82 (5), 1751-1758

      Two miniature mass spectrometer systems based on a microengineered quadrupole mass filter have been developed. One of the instruments has a footprint of 27 cm × 20 cm and is intended for laboratory use when space is at a premium. The other is portable and ...

  • Cover Image

    Paul Trap Mass Analyzer Consisting of Opposing Microfabricated Electrode Plates

    Zhiping Zhang, Ying Peng, Brett J. Hansen, Ivan W. Miller, Miao Wang, Milton L. Lee, Aaron R. Hawkins and Daniel E. Austin
    Analytical Chemistry2009 81 (13), 5241-5248
    • Paul Trap Mass Analyzer Consisting of Opposing Microfabricated Electrode Plates

      Zhiping Zhang, Ying Peng, Brett J. Hansen, Ivan W. Miller, Miao Wang, Milton L. Lee, Aaron R. Hawkins and Daniel E. Austin
      Analytical Chemistry2009 81 (13), 5241-5248

      We report the design and performance of a novel radio-frequency (RF) ion-trap mass analyzer, the planar Paul trap, in which a quadrupolar potential distribution is made between two electrode plates. Each plate consists of a series of concentric, ...

  • Cover Image

    Ion Trap Array Mass Analyzer: Structure and Performance

    Xiaoxu Li, Gongyu Jiang, Chan Luo, Fuxing Xu, Yuanyuan Wang, Li Ding and Chuan-Fan Ding
    Analytical Chemistry2009 81 (12), 4840-4846
    • Ion Trap Array Mass Analyzer: Structure and Performance

      Xiaoxu Li, Gongyu Jiang, Chan Luo, Fuxing Xu, Yuanyuan Wang, Li Ding and Chuan-Fan Ding
      Analytical Chemistry2009 81 (12), 4840-4846

      An ion trap array (ITA) mass analyzer—a novel ion trap mass analyzer with multiple ion trapping and analyzing channels—was designed and constructed. Its property and performance were investigated and reported in this paper. The ITA was built with several ...

Tools

SciFinder Links

SciFinder subscribers:  Click to sign in | Not a SciFinder subscriber? Learn more at www.cas.org

Explore by:


History

  • Published In Issue August 15, 2004
  • Received for review April 17, 2004. Accepted June 23, 2004.

Recommend & Share

  • Share on ACS NetworkACS Network
  • Add to FacebookFacebook
  • Tweet ThisTweet This
  • Add to CiteULikeCiteULike
  • Add to NewsvineNewsvine
  • Digg ThisDigg This
  • Add to DeliciousDelicious

Related Content

Other ACS content by these authors: