Measurement of Ammonia in Human Breath with a Liquid-Film Conductivity Sensor

Kei Toda, Jianzhong Li, and Purnendu K. Dasgupta*
Department of Chemistry and Biochemistry, Texas Tech University, Lubbock, Texas 79409-1061
Anal. Chem., 2006, 78 (20), pp 7284–7291
DOI: 10.1021/ac060965m
Publication Date (Web): September 15, 2006
Copyright © 2006 American Chemical Society

 Permanent address:  Department of Science, Kumamoto University, Kurokami, Kumamoto 860-8555, Japan.

,
*

 Corresponding author. Present address:  Department of Chemistry and Biochemistry, University of Texas at Arlington, Arlington, TX 76019-0065. E-mail:  Dasgupta@uta.edu.

Abstract

Measurement of breath NH3 is of interest in clinical applications as it can be used as a measure of kidney/liver functions as well as halitosis. We have developed a liquid-film conductivity sensor to measure NH3 in human breath. A film of dilute H2SO4 is formed on the top of two metal capillary tubes placed in a concentric annular arrangement. The tube exterior has been specially treated to render it hydrophilic. As breath passes over the sensor tip, the film collects NH3 and the solution conductivity (measured by the concentric capillaries functioning as electrodes) decreases accordingly. This initial rate of conductivity decrease was determined to be the best metric (most rapid and least dependent on breath pCO2) for ammonia, relative to time to attain complete neutralization (conductivity minimum) or the final rate of conductivity increase as more ammonia dissolves after neutralization. The absorbing solution composition was optimized so that CO2 does not interfere. Both dynamic measurement using mask sampling and offline balloon sampling were performed. Ammonia readily absorbs on surfaces when significant concentrations of water vapor are present. As such, memory effects are common when analyzing human breath for ammonia. This problem was successfully eliminated. The results from this sensor agreed well with data obtained by a solution-phase fluorometric technique using a porous membrane diffusion scrubber and o-phthalaldehyde derivatization chemistry. For breath CO2 measurement, the applicability of a similar sensor that relies on a NaOH film was also demonstrated.

Tools

History

  • Published In Issue October 15, 2006
  • Received for review May 25, 2006. Accepted August 18, 2006.

Recommend & Share

Related Content

Other ACS content by these authors: