Prion Protein Detection Using Nanomechanical Resonator Arrays and Secondary Mass Labeling

Madhukar Varshney, Philip S. Waggoner, Christine P. Tan, Keith Aubin, Richard A. Montagna,§ and Harold G. Craighead*
School of Applied and Engineering Physics, Department of Biomedical Engineering, Cornell University, Ithaca, New York 14853, and Innovative Biotechnologies International, Incorporated, Grand Island, New York 14072
Anal. Chem., 2008, 80 (6), pp 2141–2148
DOI: 10.1021/ac702153p
Publication Date (Web): February 14, 2008
Copyright © 2008 American Chemical Society

Abstract

Nanomechanical resonators have shown potential application for mass sensing and have been used to detect a variety of biomolecules. In this study, a dynamic resonance-based technique was used to detect prion proteins (PrP), which in conformationally altered forms are known to cause neurodegenerative diseases in animals as well as humans. Antibodies and nanoparticles were used as mass labels to increase the mass shift and thus amplify the frequency shift signal used in PrP detection. A sandwich assay was used to immobilize PrP between two monoclonal antibodies, one of which was conjugated to the resonator's surface while the other was either used alone or linked to the nanoparticles as a mass label. Without additional mass labeling, PrP was not detected at concentrations below 20 μg/mL. In the presence of secondary antibodies the analytical sensitivity was improved to 2 μg/mL. With the use of functionalized nanoparticles, the sensitivity improved an additional 3 orders of magnitude to 2 ng/mL.

Tools

SciFinder Links

SciFinder subscribers:  Click to sign in | Not a SciFinder subscriber? Learn more at www.cas.org

Explore by:


History

  • Published In Issue March 15, 2008
  • Received for review October 18, 2007. Accepted January 7, 2008.

Recommend & Share

Related Content

Other ACS content by these authors: