Table 1. Interferences in Atomic Spectroscopy

Technique Type of Interference Method of Compensation
Flame AAS Ionization
Chemical
Physical
Ionization buffers
Releasing agent or nitrous oxide-acetylene flame
Dilution, matrix matching, or method of additions  
GFAAS Physical and chemical
Molecular absorption
Spectral
Standard Temperature Platform Furnace (STPF) conditions, standard additions
Zeeman or continuum source background correction
Zeeman background correction
ICP-OES Spectral
Matrix
Background correction or the use of alternate analytical lines
Internal standardization
ICP-MS Spectral

Matrix

Interelement correction, use of alternate masses, higher resolution systems or reaction/collision cell technology
Internal standardization